Specifications
NICO® 2 Respiratory Profile Monitor
Ventilaton Modes

Adult patients on controlled mechanicalventilation, or patients breathing spontaneously with ventilation support.

Ventilaton Modes

Partial CO2 rebreathing differential FICK, noninvasive.

Measurement Frequency

Rebreathing measurementmade every three minutes for a duration of 35 seconds.

Nico Sensor
Valve type Dual diaphragm, pneumatically controlled
Return Spring 4 to 40 cm H2O
Rebreathed Volume Normal position: 35 ml
Rebreathing Position:
  • 125-285 ml (small)
  • 150-450 ml (standard)
  • 200-835 ml (large)
CO2 / Flow Sensor Integrated into NICO Sensor
Alerts
Adjustable Limits: C.O., ETCO2, SpO2, RR, No Respiration, VCO2 and Pulse rate
Audio: Adjustable volume, 2 min. silence or OFF (LED indicator)
Visual: On-screen indication and red LCD for high-priority alerts.
Internal Battery
Life: 45 minutes on full-charge
Recharge Time : 12 hours
Type: Lead acid gel cell

Top Of Page